Oral Maxillofac Surg ; 2024 Feb 06.
Artigo em Inglês | MEDLINE | ID: mdl-38319397

RESUMO

PURPOSE: Jaw-stretching devices, including the Amplification, Resistance, and Kinetics of the Jaw (ARK-JSD), are an effective option for treating trismus after head and neck cancer (HNC) treatment. The force, however, that is applied to the patient's jaw is unknown. METHODS: Ten ARK-JSD devices were constructed for each of the levels of resistance (total of 30 samples). Each sample was tested using a Universal Testing Machine (UTM). RESULTS: The easy, medium, and hard ARK-JSD had a mean maximum force of 12.3, 21.0, and 32.7 Newtons (N) at a mean interincisal distance (IID) of 8.0 mm, 13.0 mm, and 16.0 mm, respectively. The force varied by 6.9 N for the easy and 24.1 N for the hard ARK-JSD. Fatigue analysis demonstrated up to 5.5 N loss of force over 10 weeks. CONCLUSION: The ARK-JSD is a low-cost trismus device that can force between 12.3 and 32.7 N. The variation in resistance may impact efficacy. Understanding this variation will assist clinicians and patients using the ARK-JSD for trismus therapy.